Hioki Automatic Insulation / Withstanding Hitester 3153

0.2 kV to 5 kV AC Programmable and Fully Remote Controllable Automatic Insulation and Hipot Tester

The Hioki electrical safety test instrument series is designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. The 3153 is an automatic insulation and AC/DC voltage withstanding tester that fully supports automated testing.

Programmable insulation (50 to 1,200 V DC) and dielectric strength (AC/DC) testing
• Program up to 32 files of test types, test points (50 steps), and measurement settings
• Optional scanner for multi-point automatic testing
• Uses the PWM method to generate accurate test voltages that do not depend on the supply voltage
• Ramp timer function for increasing or decreasing the applied voltage during dielectric strength testing at user-specified times

■ Basic specifications (Accuracy guaranteed for 1 year)

Functions Program up to 32 files of 50 step test settings.
10 sets each of dielectric strength and insulation test settings, hold, buzzer
Monitor functions Output voltage, detected current, measured resistance, Refresh rate: 2 times/s
Power supply 100 to 120 V, 200 to 240 V AC, (50/60 Hz), 1000 VA max.
Dimensions and mass 320 mm (12.60 in) W × 155 mm (6.10 in) H × 480 mm (18.9 in) D, 18 kg (634.9 oz)
Accessories H.V. Test Lead 9615 (high voltage side and return, 1 each) ×1, Power cord ×1, Instruction manual ×1, Spare fuse ×1

[Withstanding test]

Testing voltage AC/DC 0.2 kV to 5.00 kV AC, 500 VA (max. 30 minutes), 0.2 kV to 5.00 kV DC, 50 VA (continuance)
Voltage setting Digital setting (0.01 kV setting resolution)
Waveform/Frequency Sine wave (5% or less distortion, unloaded), 50/60 Hz selectable
Current measurement 0.01 mA to 100.0 mA, Average rectified display (Digital)
Measurement range 10 mA (0.01 mA resolution), 100 mA (0.1 mA resolution)
Voltmeter Digital: accuracy ±1.5 % f.s. (f.s.=5.00 kV) (Average rectified display)
Decision method Window comparison (digital settings)

[Insulation test]

Rated testing voltage 50 to 1,200 V DC (in 1 V steps)
Rated testing current 1 mA, Short-circuit current: 200 mA or less
Measurement range/accuracy 0.10 to 9999 MΩ, 4 ranges, ± 4 % rdg. (representative values for 0.5 MΩ to 1,000 MΩ)
Decision method Window comparison (digital settings)

[Timer]

*Test times may differ from set timer times depending on the load.

Setting range 0.3 to 999 s
Ramp/delay Testing voltage ramp-up, or down, Insulation test delay: 0.1 to 99.9 s